Answers are grounded in this paper and cite their sources.
Objectives Paclitaxel-coated balloon angioplasty has been established as the first-line therapy for femoropopliteal artery disease. The primary objectives of the study were to evaluate the performance and the safety of the GORE-DCB Catheter in the treatment of atherosclerotic femoropopliteal lesions in patients with peripheral artery disease for CE-Mark approval.Methods Prospective, single-arm, multicenter study with 24 months follow-up. The GORE-DCB Catheter consists of a drug-coated nylon (inner layer)/ePTFE (outer layer) composite balloon. The ePTFE layer is coated with paclitaxel (concentration: 3.5 μg/mm2) and the excipient stearic acid/tromethamine (tris). The primary endpoints were 6-month late lumen loss (LLL) and 30-day of freedom from Major Adverse Events (MAE).Results Fifty-two subjects were enrolled, 69% men, median age 69 (49–83) years. Acute device success was 100%, the 30-day MAE rate was zero. Study primary endpoint of LLL (−0.17 mm) showed significant superiority compared to the performance goal of uncoated PTA balloon catheters from literature. At 1 and 2 years, primary patency rates were 81.8% and 68.7%, respectively, and freedom from clinically driven target lesion revascularization rates was 87.9% and 83.4%, respectively.Conclusion The study demonstrates that the treatment of lesions in femoropopliteal arteries with the GORE-DCB Catheter is safe and effective.
No takes yet. Share an insight, caveat, or question.
Noory et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: