Answers are grounded in this paper and cite their sources.
We describe and characterize a dual-channel incoherent broadband cavity-enhanced absorption spectrometer (DC-IBBCEAS) for the sensitive measurements of NOx: the sum of nitrogen monoxide (NO) and nitrogen dioxide (NO2) in the atmosphere. The instrument employs two 1 m long optical cavities, with air being extracted from a common sampling line. The first channel (cavity-1) spans 340-380 nm and the second channel (cavity-2) spans the 405-460 nm spectral range, both measuring NO2 simultaneously. High absorption cross-sections of NO2 in both channels are effectively utilized for its sensitive quantification. NO is quantified by titrating it with ozone to NO2 in channel-2, where the difference of NO2 measured from that in channel-1 corresponds to the NO concentration in the sampled air. The instrument offers 1-ppb detection sensitivity for both NO and NO2 with a maximum possible uncertainty of ∼9%. The use of close yet different spectral regions in the two channels readily extended measurements to a broader range without compromising its sensitivity to NOx quantification. This would extend the DC-IBBCEAS applicability to simultaneously monitor interfering species with significant absorption cross-sections in the region in either channel, such as glyoxal (CHOCHO), methylglyoxal (CH3COCHO), and nitrous acid (HONO).
No takes yet. Share an insight, caveat, or question.
Pakkattil et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: