This is the authors' abstract. We don't add key points for this paper.
We present an all-optical photoacoustic (PA) multigas analyzer by using a digital fiber-optic acoustic detector, which consists of a spectrum demodulation based fiber Fabry-Perot (F-P) pressure sensor and a digital virtual lock-in amplifier. The concentration of multigas is quantified by using an intensity-modulated infrared thermal source and two wavelength-modulated laser diodes. Light beams from the three sources are coupled to a single nonresonant PA cell. The generated PA pressure is detected by the fiber F-P pressure sensor, which is demodulated by a near-infrared spectrometer. A lock-in white-light interferometer is developed for detecting first-harmonic (1f ) and second-harmonic (2f ) signals. The noise equivalent detection limits with a 100-s averaging time are 37, 9, 6, 17, 4, and 60 ppb for CH4, C2H2, C2H4, C2H6, CO, and CO2, respectively.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Chen et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: