Heavyweight pacemaker pulse generators can cause late displacement of endocardial electrodes due to downward migration, highlighting the need for proper anchoring or subpectoral pocket creation.
Answers are grounded in this paper and cite their sources.
Heavy generators may cause late lead retraction; this case leaves open optimal anchoring or pocket techniques.
ENDOCARDIAL pacing has been shown to be a satisfactory method for long-term cardiac pacing. The electrode complications of malposition, displacement, fracture, migration, and perforation have been well described in previous publications.1,2We have recently observed a retraction of the endocardial electrode by the downward migration of the heavy pulse generator. This complication is prevented by anchoring the pacemaker to the clavicle or by creating a pacemaker "pocket" under the pectoral muscle or both. Report of Cases A 78-year-old man was admitted to New York University Hospital with intermittent heart block requiring cardiac pacing. A Medtronic pacemaker weighing 200 gm was inserted in a subcutaneous infrapectoral pocket with a bipolar electrode inserted via the cephalic vein into the apex of the right ventricle. Postoperatively the patient did well until the 12th month when failure of capture was noted. Chest x-ray film showed retraction of the catheter into the right atrium