Key result
A sheath-retracting active-fixation pacemaker electrode dislodged and could not be repositioned because impacted endocardial tissue prevented the screw from being re-advanced out of its housing.
Case Report (n=1)
Answers are grounded in this paper and cite their sources.
Impacted endocardium can prevent the repositioning of a sheath-retracting active-fixation pacemaker electrode by blocking the screw from re-advancing.
A patient is described in whom a sheath‐retracting active‐fixation pacemaker electrode dislodged shortly after successful transvenous implantation in the right atrial appendage. Attempts to reposition the lead were unsuccessful. Examination of the explanted electrode disclosed that inability to reposition the electrode was due to the fact that impacted tissue around the electrode “screw” prevented its being re‐advanced out of its polyurethane housing. (PACE, Vol. 5, March‐April, 1982)
No takes yet. Share an insight, caveat, or question.
Salem et al. (1982) conducted a case report in Pacemaker electrode dislodgement (n=1). Sheath-retracting active-fixation pacemaker electrode was evaluated on Electrode malfunction and inability to reposition. A sheath-retracting active-fixation pacemaker electrode dislodged and could not be repositioned because impacted endocardial tissue prevented the screw from being re-advanced out of its housing.