Why the study?
No published data had evaluated the exercise performance of the AVEIR VR leadless pacemaker or the relationship between intrinsic sinus heart rate and the delivered ventricular rate during exercise testing.
Population
13 patients implanted with an AVEIR VR leadless pacemaker without atrial fibrillation
Comparison
Intrinsic sinus heart rate vs paced or projected paced ventricular heart rate during bicycle ergometer exercise
Answers are grounded in this paper and cite their sources.
Initial observations on AVEIR VR rate response during exercise warrant confirmation; leaves open whether it reliably tracks sinus rates in practice.
BACKGROUND: To date, no published data have evaluated the exercise performance of the AVEIR VR leadless pacemaker. OBJECTIVE: We aimed to assess, during bicycle ergometer exercise testing, the relationship between the intrinsic sinus heart rate and the ventricular rate delivered by the AVEIR VR pacemaker. Patients with atrial fibrillation or those unable to perform exercise testing were excluded. METHODS: Patients underwent a bicycle ergometer exercise test using a workload protocol normalized to body weight (W/kg) according to their level of physical activity (sedentary or active). Simultaneous device interrogation was performed throughout the test. Intrinsic sinus heart rate and paced heart rate (or the projected paced heart rate in non-pacemaker-dependent patients) were recorded. Sensor latency was defined as the time required to achieve a 10-beats/min increase in paced heart rate above baseline. RESULTS: Between January and October 2025, 13 patients were included (69.9 ± 12.7 years; 23% women). Five patients were physically active, and eight patients were pacemaker-dependent. The mean sensor latency was 3.5 min in pacemaker-dependent patients and 2.7 min in non-pacemaker-dependent patients. Rate-response gain required adjustment in 38.5% of patients following exercise testing. Base rate and/or the maximum heart rate response were modified in 76.9% of patients. CONCLUSION: Exercise testing in patients implanted with an AVEIR VR leadless pacemaker demonstrated an overall appropriate rate-response, although with a non-negligible sensor latency. A substantial proportion of patients required post-exercise optimization of rate-response programming, highlighting the value of exercise-guided device programming to improve physiological chronotropic adaptation.
No takes yet. Share an insight, caveat, or question.
Gallet et al. (2026) studied this question.