Endocardial atrial pacing lead bodies can migrate late after placement in Fontan patients, potentially causing severe blood flow obstruction and cardiac decompensation.
Answers are grounded in this paper and cite their sources.
Raises awareness for late lead migration in Fontan patients with decompensation; leaves open optimal surveillance strategies.
Migration of intracardiac transvenous pacing leads may occur. There is a known risk of intrapulmonary ventricular pacing lead migration in patients with endocardial lead systems. In the current report we present the late intrapulmonary migration of an endocardial atrial pacing lead body. The patient had undergone antitachycardia pacemaker placement to control recurrent atrial tachyarrhythmias following the Fontan procedure. Although the lead electrode remained in place and continued to pace, the lead body migrated, causing severe obstruction to blood flow. This resulted in severe cardiac decompensation, which was ultimately ameliorated by lead repositioning.
No takes yet. Share an insight, caveat, or question.
O’Connor et al. (1996) studied this question.