Why the study?
Is a nominal ventricular output setting of 3.5 V sufficient to provide a 100% safety margin in patients undergoing pacemaker pulse generator exchange?
Is a nominal ventricular output setting of 3.5 V sufficient to provide a 100% safety margin in patients undergoing pacemaker pulse generator exchange?
A nominal ventricular output setting of 3.5 V is sufficient to provide a 100% safety margin in the vast majority of patients undergoing pacemaker pulse generator exchange.
Answers are grounded in this paper and cite their sources.
May support nominal 3.5-V output after most pacemaker generator exchanges; leaves open randomized confirmation of safety margins.
Projected pacemaker longevity is calculated according to the nominal setting, which is 3.5 V for pulse amplitude in most present day pacemakers. The aim of this study was to test whether the nominal ventricular output setting of 3.5 V and 0.4 ms provides a 100% safety margin if these pacemakers are connected to implanted ventricular pacing leads. The study included 24 patients undergoing pulse generator exchange. The new pacemaker was either a Thera DR (n = 21) or Thera SR (n = 3) device. Ventricular pacing leads were implanted 70 +/- 38 months previously. Intraoperative measurements included pacing threshold at 0.5-ms pulse duration, impedance, and R wave amplitude. To achieve a 100% safety margin with 3.5 V, the cut-off pacing threshold is 1.7 V. At discharge it was assessed whether ventricular pulse amplitude remained at < or = 3.5 V or was programmed to > 3.5 V. At pulse generator exchange, pacing threshold was 1.2 +/- 0.5 V, including four patients with pacing thresholds > 1.7 V. Impedance was 587 +/- 189 omega and R wave amplitude was 12.8 +/- 4.8 mV. At discharge, ventricular pulse amplitude remained at the nominal setting in 13 patients, including 2 patients with high pacing thresholds at pacemaker exchange, and was programmed to < 3.5 V in 7 patients. Ventricular pulse amplitude was programmed to > 3.5 V in four patients. Two of these patients had high pacing thresholds (> 1.7 V/0.5 ms) at pulse generator exchange; the other two patients were programmed to > 3.5 V, although 3.5 V already provided a 100% safety margin. After pulse generator exchange, 92% of the patients could be paced with a 3.5-V pulse amplitude. Pacemakers were programmed in four patients > 3.5 V, but in only two of them to obtain a sufficient safety margin. High pacing thresholds at pulse generator exchange did not generally predict high pacing thresholds at discharge.
No takes yet. Share an insight, caveat, or question.
Schuchert et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: