Iomeprol administration during angiocardiography can cause laboratory interference on capillary zone electrophoresis, mimicking a monoclonal protein.
Answers are grounded in this paper and cite their sources.
Clinicians should consider iomeprol interference when beta-fraction peaks appear post-angiography; this case extends the catalog of known electrophoresis artifacts.
Analysis of serum proteins by capillary zone electrophoresis (CZE) has become a routine method for the detection of dysproteinemias and monoclonal proteins. Detection is based on ultraviolet absorbance at 200 nm (Capillarys, Sebia) or 214 nm (Paragon 2000, Beckman Coulter), which corresponds to the absorption of peptide bonds in proteins. Compared with gel electrophoresis, CZE offers the advantage that proteins are more proportionately quantified compared with the variable dye-staining intensities exhibited by different serum proteins. A drawback of CZE, however, is the potential interference by exogenous nonprotein substances, such as radioopaque agents and antibiotics, which also absorb at these wavelengths (1). A 5-year old girl with protein-losing enteropathy underwent routine angiocardiography 4 years after a Fontan operation for a univentricular heart, and received 6–7 mL/kg iomeprol (Iomeron). A venous blood sample was sent to the laboratory for serum protein electrophoresis. The CZE with the Capillarys showed hypogammaglobulinemia and a suspect peak in the β-fraction (Fig. 1A1 ). Subsequent immunofixation (Sebia Hydrasys), however, failed to demonstrate a monoclonal protein (Fig. 1A inset). Consultation with the clinicians revealed that the blood sample had been drawn via a venous access during the angiocardiography procedure, raising the suspicion that the peak could be a result of interference from the iodinated contrast agent.
No takes yet. Share an insight, caveat, or question.
Vermeersch et al. (2006) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: