Why the study?
Does intracardiac oximetry using a fiberoptic reflection oximeter improve diagnostic capabilities compared to standard sampling in patients with suspected congenital heart disease?
Does intracardiac oximetry using a fiberoptic reflection oximeter improve diagnostic capabilities compared to standard sampling in patients with suspected congenital heart disease?
The use of a fiberoptic reflection oximeter during cardiac catheterization in congenital heart disease provides immediate, precise oxygen saturation data without blood loss compared to standard sampling.
Answers are grounded in this paper and cite their sources.
May enable real-time, bloodless oximetry during congenital catheterization; hypothesis-generating and should not yet change practice.
A fiberoptic reflection oximeter has been adapted to the diagnostic study of patients suspected of having congenital heart disease. The results of 10 successive catheterizations with use of this technic are reported and compared with the results obtained in the same subjects with a standard sampling technic. A large number of determinations was possible without blood loss. The results were immediately available for use in directing the manipulation of the catheter. Freedom from the distortions attendant upon catheter sampling permitted precise observation of the nature of flow across a ventricular septal defect and of the relationship between blood oxygen saturation in the main pulmonary artery and the pulmonary arterial wedge position.
No takes yet. Share an insight, caveat, or question.
Enson et al. (1964) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: