Why the study?
CIED implantation is increasing in low- and middle-income countries, but data on CIED removal experience and outcomes from resource-constrained settings were needed.
Percutaneous transvenous CIED lead removal in a resource-constrained South African center is feasible with high success and low complication rates, though long-term mortality remains high due to underlying patient comorbidities.
BACKGROUND: The rate of cardiac implantable electronic device (CIED) implantation in low- and middle-income countries is increasing. Patients recieving these devices are frequently older and with multiple co-morbidities, which may later lead to complications requiring CIED removal. CIED removals are associated with life-threatening complications. However, high sucesss rates are reported in high-income countries. The purpose of this study was to report on the experience of CIED removal in a resource-constrained setting. METHODS: In this retrospective study, we included consecutive adult patients admitted to Groote Schuur Hospital and the University of Cape Town Private Academic Hospital for CIED removal from 1 January 2008 to 31 December 2019. RESULTS: During the study period, 53 patients underwent CIED removal (26 extractions and 27 explants). The patients had a mean (standard deviation) age of 59.1 (16.0) years. A history of systemic hypertension was present in 50.9% of patients, diabetes mellitus in 30.2% and dilated cardiomyopathy in 47.2%. Complete heart block was the leading indication for CIED implantation (37.7%), and device infection was the leading indication for removal (69.2%). CIEDs were removed after a median (interquantile range) of 243 (53-831) days. There were 40 leads extracted and 35 explants. Lead extractions were perfomed in the cardiac catheterisation laboratory under general anaesthesia via a percutaneous transvenous superior approach. There was one major and one minor complication related to lead extraction. CONCLUSIONS: CIED infections were the primary indication for CIED removal in a tertiary referral centre in South Africa. Despite being a low-volume centre, we report a high percutaneous transvenous extraction success rate with low complication rate; results which are comparable to high-volume centres.
No takes yet. Share an insight, caveat, or question.
Mkoko et al. (2021) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: