Answers are grounded in this paper and cite their sources.
Currently, patients have to keep track of doses to determine when to replace their metered-dose inhalers (MDIs). This study evaluated the performance and patient satisfaction of a novel MDI with an integrated dose counter. In an open-label study at 38 outpatient centres, patients > or =12 years old with asthma or chronic obstructive pulmonary disease (COPD) received two actuations of fluticasone propionate/salmeterol 125/25 microg (115/21 microg ex-actuator) hydrofluoroalkane (ADVAIR) HFA) via MDI with counter twice a day until all 120 actuations were completed. Concordance between counter and diary recordings in patients who reported use of > or =90% of labelled actuations (completer population, n = 228) was high (discrepancy rate of 0.94%) and the incidence of device firing without changes in counter readings was low (0.13%). Mean expected actuations based on canister weights (114) were slightly lower than mean counter (121) and diary reported actuations (120). Upon study completion, 95% of patients were satisfied with the dose counter and 92% agreed it would help prevent them from running out of medication. Safety assessments (intent-to-treat population, n = 237) indicated that the drug was well tolerated. This integrated MDI counter may help patients maintain better disease control by enabling them to accurately track their medication supply.
No takes yet. Share an insight, caveat, or question.
Sheth et al. (2006) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: