The study demonstrates a technical method using Doppler and optical fibers to measure blood flow by filtering out involuntary body movements.
Answers are grounded in this paper and cite their sources.
Feasibility of movement-filtered optical Doppler blood flow measurement shown; leaves open clinical validation and utility in cardiovascular monitoring.
Skin measurements made using the Doppler method can be used to monitor blood flow and involuntary body movements. The spectra of the Doppler signal indicate that it is dominated by the involuntary body movements and that the blood flow is detectable only at higher frequencies. Some separation of the two signals is therefore possible if an active high-pass filter is used. By this means the blood-flow parameter has been measured during artificial variations of the blood flow and at different parts of the body. A special arrangement of optical fibres allows the scattered light and its mean intensity at the same location to be detected separately. The influence of intensity variations is compensated without using any time constant.
No takes yet. Share an insight, caveat, or question.
Ruth et al. (1984) studied this question.