BACKGROUND/AIM: This study aimed to compare the use of a rotating gantry in liver tumor carbon-ion radiotherapy using of a fixed-port for treatment planning. MATERIALS AND METHODS: Thirty patients with liver tumors were analyzed. Three treatment plans were developed for each case: one with a rotating gantry with a 360° angle, one with fixed ports of 0° and 90° with a ±20° couch rolling setting, and one with fixed ports of 45° and 90° with a ±20° couch rolling setting. The dose-volume histogram parameters of the clinical target volume (CTV) and organs at risk (OARs) for each treatment plan were compared. RESULTS: Significant differences in the volume of the liver-gross tumor volume (GTV) of normal liver irradiated with 5 Gy to 15 Gy were found between the gantry treatment plans and fixed-port treatment plans. There were no significant differences in the OARs, except for the CTV and liver GTV, between the gantry and fixed-port treatment plans. CONCLUSION: The study results support the potential of using a rotating gantry to reduce liver doses, especially in the low-to-medium dose range, while maintaining target and OAR doses except for the liver. A rotating gantry could be especially useful in cases in which the relationship between the tumor and OAR is complicated by location.