Why the study?
Do the Del Mar Avionics Pressurometer IV and Spacelabs 90202 ambulatory blood pressure monitors provide accurate readings compared to trained observers in normotensive and hypertensive subjects?
Do the Del Mar Avionics Pressurometer IV and Spacelabs 90202 ambulatory blood pressure monitors provide accurate readings compared to trained observers in normotensive and hypertensive subjects?
Both the Del Mar Avionics Pressurometer IV and Spacelabs 90202 are satisfactory and reliable for ambulatory BP monitoring, with slight biases compared to auscultatory readings.
Answers are grounded in this paper and cite their sources.
Acceptable agreement supports reliability of these ambulatory monitors; leaves open need for prospective validation in diverse populations.
The purpose of this study was to evaluate the accuracy of two noninvasive ambulatory blood pressure (BP) monitors, the new Del Mar Avionics Pressurometer IV (PIV) and the second generation Spacelabs 90202. Two sets of comparisons were made between two trained observers and the monitors on 17 normotensive and 28 hypertensive subjects. The result of the first comparison showed that the average difference between observers and the PIV was 1.2/-2.2 mm Hg for systolic and diastolic BP. Comparisons of both monitors to the observers showed differences of 0.9/1.0 mm Hg for the PIV and 0.3/0.8 mm Hg for the 90202 (systolic/diastolic). Correlations between the average observer reading and the monitors over both sets of comparisons ranged from 0.90 (diastolic 90202 vs observer) to 0.98 (systolic PIV vs observer). The percentage of readings within 5 mm Hg between observer and monitor over all comparisons ranged from a low of 68% (90202, systolic and diastolic) to a high of 83% (PIV, diastolic). In general, both devices are satisfactory and reliable machines for ambulatory BP monitoring, but each has a slightly different bias compared to auscultatory readings.
No takes yet. Share an insight, caveat, or question.
Santucci et al. (1989) studied this question.