Answers are grounded in this paper and cite their sources.
This work describes a novel electrochemical sensor fabricated by an injection molding process. This device features a conductive polymer electrode encased in a plastic holder and electroplated in situ with a thin antimony film. The antimony film sensor was applied to the determination of Pb(II) and Cd(II) by anodic stripping voltammetry (ASV). The deposition of Sb on the sensor was studied by cyclic voltammetry (CV) and microscopy. The experimental variables (concentration of the antimony plating solution, deposition potential and time, stripping waveform) were investigated, and the potential interferences were studied and addressed. The limits of detection were 0.95 μg L−1 for Pb(II) and 1.3 for Cd(II) (at 240 s of preconcentration) and the within-sensor percentage relative standard deviations were 4.2% and 4.9%, respectively, at the 25 μg L−1 level (n = 8). Finally, the sensor was applied to the determination of Pb(II) and Cd(II) in a phosphorite sample and a lake water sample.