Key result
Olink-measured ET-1 positively correlates with ELISA-derived levels in coronary artery plasma.
Why the study?
It was unknown whether the Olink Explore Proximity Extension Assay platform specifically measures biologically active ET-1 or inactive EDN1-encoded peptides.
Does the Olink Explore platform accurately measure biologically active ET-1 compared to an ET-1 specific ELISA in patients with coronary artery disease?
Observational (n=29)
Does the Olink Explore platform accurately measure biologically active ET-1 compared to an ET-1 specific ELISA in patients with coronary artery disease?
The Olink Explore platform can detect relative changes in biologically active ET-1, supporting its use as a biomarker tool in clinical and translational studies.
Olink ET-1 assay correlates with ELISA in coronary plasma; supports biomarker use in CAD studies but leaves open need for larger validation.
Endothelin-1 (ET-1) antagonists are increasingly being approved for new treatments for cardiovascular disease, where elevated ET-1 levels contribute to increased vasoconstriction. Further therapeutic targets, including coronary artery disease, are under investigation. The Olink Explore 3072 Proximity Extension Assay platform enables multiplexed high throughput measurement of ~3000 plasma proteins, from minimal (≤6µl) sample volumes. However, it is not known if the two oligonucleotide‑tagged antibodies raised against preproET-1₁-₂₁₂, used in this Olink assay, specifically measure biologically active ET-1 or the other inactive EDN1-encoded peptides, also secreted by human endothelial cells. Paired plasma samples from 29 patients with coronary artery disease were obtained, using a specialised intra-coronary sampling catheter, designed to obtain site specific biochemical information from within coronary arteries. We compared ET-1 concentrations measured with an ET-1 specific ELISA, demonstrated to have no cross reactivity with other EDN1-encoded peptides versus values obtained using Olink Explore platform. Olink-measured ET-1 correlated significantly with ELISA-derived ET-1 levels (r=0.53, p=0.003), and Olink values predicted ELISA results. Olink ET-1 concentrations also correlated with ETB receptor levels (r=0.40, p< 0.05). These findings indicate that the Olink Explore platform can detect relative changes in biologically active ET-1, supporting its use as a biomarker tool in clinical and translational studies.
No takes yet. Share an insight, caveat, or question.
Anwar et al. (2026) conducted an observational in Coronary artery disease (n=29). Olink Explore platform vs. ET-1 specific ELISA was evaluated on Correlation of ET-1 concentrations between Olink Explore platform and ET-1 specific ELISA (r=0.53, p=0.003). Olink-measured ET-1 concentrations correlated significantly with ELISA-derived ET-1 levels in human plasma from coronary arteries (r=0.53, p=0.003).