A Varian 4-MV isocentric therapy accelerator has been modified to perform also as a CT scanner. The goal is to provide a computed tomography capability for use in radiotherapy at low cost. We envision three principal uses for the system. These are (i) to provide two and three-dimensional maps of electron density distribution for CT assisted therapy planning; (ii) to aid in patient set up by providing sectional views of the treatment volume and high contrast scout-mode verification images; and (iii) to provide a means for periodically checking the patients anatomical conformation against that which was used to generate the original therapy plan. The machine was modified by mounting an array of detectors on a frame that is bolted to the counterweight end of the gantry in such a manner as to define a "third generation" CT scanner geometry. Details of the detectors and data acquisition electronics are provided. We present results of phantom studies which demonstrate that this prototype system has a spatial resolution of 4 mm and the ability to discriminate electron density differences of less than 1%. We also show that the system is exactly linear over a wide range of electron densities corresponding to those found in body tissues.