Key result
CoreValve transcatheter aortic valve implantation significantly prolonged QRS, PR, AH, and HV intervals, causing persistent conduction disturbances in 78% of patients.
Why the study?
Does CoreValve transcatheter aortic valve implantation affect atrioventricular conduction in pacemaker-free patients?
Observational (n=18)
Does CoreValve transcatheter aortic valve implantation affect atrioventricular conduction in pacemaker-free patients?
CoreValve TAVI frequently causes direct damage to the His bundle or AV node, leading to transient or permanent AV conduction worsening in the majority of patients.
Answers are grounded in this paper and cite their sources.
CoreValve TAVI warrants close post-procedural conduction monitoring; leaves open questions on long-term pacing needs and valve-specific risks.
BACKGROUND: Atrioventricular (AV) block is one of the most frequent complications of CoreValve transcatheter aortic valve implantation (TAVI). The aim of this study was to analyze the effects of CoreValve implantation on AV conduction. METHODS AND RESULTS: Electrophysiological study was performed immediately before and after CoreValve implantation in 18 consecutive, permanent pacemaker-free patients. An electrode was placed on the His bundle during valve implantation, and data were continuously recorded during the procedure. With surface ECG, a median (first, third quartile) QRS width of 96 (84, 116) to 150 (121, 164) ms (P=0.001) and PR interval of 180 (159, 216) to 210 (190, 240) ms (P=0.008) were significantly prolonged, and QRS axis was left deviated 30° (-32°, 46°) to -20° (-60°, 2°) (P=0.005). With intracardiac electrograms, the AH (97 [70, 123] to 115 [96, 135] ms, P=0.021) and HV (52 [42, 55] to 60 [50, 70] ms, P=0.002) intervals were increased. At the end of the procedure, we observed significant ECG- or electrophysiological study-persistent conduction disturbances in 14 (78%) patients. Five patients experienced transient changes (2 AV blocks and 3 left bundle branch blocks). CONCLUSIONS: CoreValve implantation worsens AV conduction in most patients, either transiently or permanently. This worsening is the result of direct damage either on the His bundle or on the AV node.
No takes yet. Share an insight, caveat, or question.
Rubín et al. (2011) conducted an observational in Atrioventricular conduction disturbance in TAVI (n=18). CoreValve transcatheter aortic valve implantation vs. Pre-implantation baseline was evaluated on ECG- or electrophysiological study-persistent conduction disturbances. CoreValve transcatheter aortic valve implantation significantly prolonged QRS, PR, AH, and HV intervals, causing persistent conduction disturbances in 78% of patients.
Synapse has enriched 10 related papers. Consider them for comparative context: