In an in vitro PVAD model, the Sorin Bicarbon bileaflet valve demonstrated faster closure and higher maximum flow velocity but generated higher maximum Reynolds shear stress compared to the Medtronic Hall monoleaflet valve.
Answers are grounded in this paper and cite their sources.
Higher shear stress with S-B valve warrants caution in PVADs; leaves open optimal mechanical valve choice pending clinical data.
Our group is developing a new type of pulsatile pneumatic ventricular assist device (PVAD) that uses the Medtronic Hall tilting disc valve (M-H valve). Although tilting disc valves have good washout effect inside the blood pump, they are no longer in common clinical use and may be difficult to obtain in the future. To investigate the stability of the Sorin Bicarbon valve (S-B valve) in our PVAD, we constructed a model pump made of an acrylic resin with the same configuration as our PVAD and attempted to compare the flow visualization upstream and downstream of the outlet position valve between the M-H valve and the S-B valve using a particle image velocimetry (PIV) method. The outlet S-B valve had faster closure than the M-H valve. The maximum flow velocity was greater than with the M-H valve. The maximum Reynolds shear stress (RSS) of the M-H valve reached 150 N/m(2) and that of the S-B valve reached 300 N/m(2) upstream during the end-systolic and early-diastolic phases. In both valves, the maximum RSS upstream of the valve was higher than downstream of the valve because of the regurgitation flow during valve closure. In addition, the maximum viscous shear stress reached above 2 N/m(2), which occupied only about 1%-1.5% of the maximum RSS.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2010) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: