Key result
Bileaflet mechanical heart valves in the Polvad directed the main jet more centrally with lower shear stress but higher turbulent stress compared to single-disc valves.
Bileaflet mechanical heart valves can successfully replace single-disc valves in the Polish ventricular assist device, demonstrating acceptable flow and stress profiles in simulations and successful initial clinical use.
Bileaflet valves offer favorable Polvad flow profiles in animal models; hypothesis-generating for VAD design and requires clinical validation.
The Polish ventricular assist device (Polvad) has been used successfully in clinical contexts for many years. The device contains two single-disc valves, one at the inlet and one at the outlet connector of the pneumatic pump. Unfortunately, in recent years, a problem has occurred with the availability of single-disc valves. This article presents the possibility of using bileaflet mechanical heart valve prostheses in the Polvad to avoid a discontinuity in clinical use. The study is based on experimental and numerical simulations and comparison of the distribution of flow, pressure, and stress (wall, shear, and turbulent) inside the Polvad chamber and the inlet/outlet connectors fitted with Sorin Monodisc and Sorin Bicarbon Fitline valves. The type and orientation of the inlet valve affects valve performance and flow distribution inside the chamber. Near-wall flow is observed for single-disc valves. In the case of bileaflet valves, the main jet is directed more centrally, with lower shear stress but higher turbulent stress in comparison with single-disc valves. For clinical usage, a 45° orientation of the bileaflet inlet valve was chosen, as this achieves good washing of the inlet area near the membrane paste surface. The Polvad with bileaflet valves has now been used successfully in our clinic for over a year and will continue to be used until new assist devices for heart support are developed.
No takes yet. Share an insight, caveat, or question.
Małota et al. (2015) studied Heart failure requiring ventricular assist device. Bileaflet mechanical heart valve prostheses (Sorin Bicarbon Fitline) vs. Single-disc valves (Sorin Monodisc) was evaluated on Distribution of flow, pressure, and stress inside the Polvad chamber. Bileaflet mechanical heart valves in the Polvad directed the main jet more centrally with lower shear stress but higher turbulent stress compared to single-disc valves.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: