Key result
Reinforced polymer bileaflet valves match Medtronic-Hall mean flow with ~1 month of durability.
A newly developed polymer bileaflet valve with 150 micrometer leaflets and reinforced seats demonstrated adequate hydrodynamics and durability for potential use in a low-cost pulsatile blood pump.
Answers are grounded in this paper and cite their sources.
May support low-cost pump development; leaves open translation of polymer valve durability to clinical use.
The final goal of this study is to realize a low-cost pulsatile blood pump especially for patients with acute heart failure or postoperative low cardiac output syndrome. In support of the pump, two types of polymer bileaflet valves with different configuration of the valve seats were developed. Influence of the leaflet thickness on the hydrodynamics of the prototype was preliminarily investigated among 70 microm, 100 microm, and 150 microm. As to the valves with the thinner leaflets, buckling of the leaflets was observed, which induced a large amount of regurgitation at valve closure. However, by thickening the leaflet to 150 microm, the mean flow of the prototype and the second model could be successfully comparable to the Medtronic-Hall valve. Moreover, accelerated fatigue tests showed that reinforcement of the valve seat with the additional spokes in the second model extended the durability by four times as compared with the prototype, equivalent to an in vivo duration of over one month.
No takes yet. Share an insight, caveat, or question.
Iwasaki et al. (2003) studied Acute heart failure or postoperative low cardiac output syndrome. Polymer bileaflet valve vs. Medtronic-Hall valve was evaluated on Hydrodynamics (mean flow) and durability. A polymer bileaflet valve with 150 microm leaflet thickness achieved mean flow comparable to the Medtronic-Hall valve, and reinforcement extended durability to an equivalent of over one month.
Synapse has enriched 10 related papers. Consider them for comparative context: