Why the study?
Does the CGCI remote magnetic navigation system allow accurate and safe catheter positioning and ablation in patients with atrial flutter?
Does the CGCI remote magnetic navigation system allow accurate and safe catheter positioning and ablation in patients with atrial flutter?
The CGCI remote magnetic navigation system demonstrates technical feasibility for accurate and stable catheter positioning during ablation of atrial flutter.
Answers are grounded in this paper and cite their sources.
Provides preliminary support for CGCI in atrial flutter ablation; hypothesis-generating and should not yet change practice.
New remote navigation systems have been developed to improve current limitations of conventional manually guided catheter ablation in complex cardiac substrates such as left atrial flutter. This protocol describes all the clinical and invasive interventional steps performed during a human electrophysiological study and ablation to assess the accuracy, safety and real-time navigation of the Catheter Guidance, Control and Imaging (CGCI) system. Patients who underwent ablation of a right or left atrium flutter substrate were included. Specifically, data from three left atrial flutter and two counterclockwise right atrial flutter procedures are shown in this report. One representative left atrial flutter procedure is shown in the movie. This system is based on eight coil-core electromagnets, which generate a dynamic magnetic field focused on the heart. Remote navigation by rapid changes (msec) in the magnetic field magnitude and a very flexible magnetized catheter allow real-time closed-loop integration and accurate, stable positioning and ablation of the arrhythmogenic substrate.
No takes yet. Share an insight, caveat, or question.
Filgueiras‐Rama et al. (2013) studied this question.