Key result
Upper limit of vulnerability >15 J during ICD implant linked to ~6.2-fold higher VF occurrence.
Why the study?
The clinical utility of the upper limit of vulnerability (ULV) test during defibrillator implantation and its association with ventricular fibrillation outcomes was evaluated.
Does an upper limit of vulnerability >15 J predict the occurrence of ventricular fibrillation in patients receiving an implantable cardioverter-defibrillator?
Cohort (n=165)
Open-label
No
Does an upper limit of vulnerability >15 J predict the occurrence of ventricular fibrillation in patients receiving an implantable cardioverter-defibrillator?
Hazard Ratio: 6.25 (95% CI 1.913–20.4)
p-value: p=<0.01
A high upper limit of vulnerability (>15 J) during ICD implantation strongly predicts future occurrences of ventricular fibrillation, indicating a link between cardiac vulnerability to electrical shock and electrical instability.
BACKGROUND The utility of the upper limit of vulnerability (ULV) test in patients undergoing defibrillator implantation has been reported, so the purpose of this study was to evaluate the difference in the clinical outcomes between patients with ULV ≤15 J or >15 J. METHODS AND RESULTS A total of 165 patients receiving an implantable cardioverter-defibrillator underwent a vulnerability test. At the time of the implantation, we delivered a 15-J shock on the T-peak and ±20 ms later to cover the most vulnerable part of the cardiac cycle. The clinical outcomes were prospectively analyzed. A 15-J shock induced ventricular fibrillation (VF) in 30 patients (ULV >15 J) and did not in 135 (ULV ≤15 J). The characteristics of the 2 groups were comparable. After a mean follow-up of 757 days, Kaplan-Meier curve analysis showed that the ULV ≤15 J group experienced less VF than the ULV >15 J group (log-rank P=0.003). The occurrence of ventricular tachycardia was similar between the 2 groups (P=0.140). Furthermore, the effectiveness of ATP was comparable. After adjusting for other known predictors of shock therapy, a ULV >15 J was independently associated with the occurrence of VF (hazard ratio: 6.25; 95% confidence interval: 1.913-20.40; P<0.01). CONCLUSIONS A high ULV value was associated with a high incidence of VF, which suggests that cardiac vulnerability to electrical shock may be linked to electrical instability.
No takes yet. Share an insight, caveat, or question.
Yamashita et al. (2014) conducted a cohort in Patients at risk of sudden cardiac death requiring ICD or CRT-D (n=165). Upper limit of vulnerability (ULV) > 15 J vs. ULV ≤ 15 J was evaluated on Occurrence of ventricular fibrillation (VF) (HR 6.25, 95% CI 1.913-20.40, p=<0.01). An upper limit of vulnerability >15 J during ICD implantation independently predicted a significantly higher occurrence of ventricular fibrillation (HR 6.25) compared to an upper limit ≤15 J.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: