Key result
Sparse cine DENSE MRI closely matches standard black blood imaging for mouse LV metrics.
Why the study?
The accuracy of magnitude images from cine displacement encoding with stimulated echoes (DENSE) MRI to quantify left ventricular volumes, mass, and ejection fraction was untested.
DENSE MRI with sparse sampling can accurately quantify LV mass, volumes, and EF in mice, significantly reducing the time required for assessment.
PurposeTo test the hypothesis that magnitude images from cine displacement encoding with stimulated echoes (DENSE) magnetic resonance imaging (MRI) can accurately quantify left ventricular (LV) volumes, mass, and ejection fraction (EF).Materials and MethodsThirteen mice (C57BL/6J) were imaged using a 7T ClinScan MRI. A short‐axis stack of cine T2‐weighted black blood (BB) images was acquired for calculation of LV volumes, mass, and EF using the gold standard sum‐of‐slices methodology. DENSE images were acquired during the same imaging session in three short‐axis (basal, mid, apical) and two long‐axis orientations. A custom surface fitting algorithm was applied to epicardial and endocardial borders from the DENSE magnitude images to calculate volumes, mass, and EF. Agreement between the DENSE‐derived measures and BB‐derived measures was assessed via coefficient of variation (CoV).Results3D surface reconstruction was completed on the order of seconds from segmented images, and required fewer slices to be segmented. Volumes, mass, and EF from DENSE‐derived surfaces matched well with BB data (CoVs ≤11%).ConclusionLV mass, volumes, and EF in mice can be quantified through sparse (five slices) sampling with DENSE. This consolidation significantly reduces the time required to assess both mass/volume‐based measures of cardiac function and advanced cardiac mechanics. J. Magn. Reson. Imaging 2014;40:398–406.
No takes yet. Share an insight, caveat, or question.
Haggerty et al. (2013) studied Evaluation of LV volumes, mass, and EF (n=13). Cine displacement encoding with stimulated echoes (DENSE) MRI vs. T2-weighted black blood (BB) images (sum-of-slices methodology) was evaluated on Agreement between DENSE-derived and BB-derived measures for LV volumes, mass, and EF (coefficient of variation) (CoVs ≤11%). Cine displacement encoding with stimulated echoes (DENSE) MRI using sparse sampling matched well with standard black blood imaging for quantifying LV volumes, mass, and EF in mice (CoVs ≤11%).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: