Does a simplified post-processing technique accurately quantify cardiac mechanics compared to the original method in humans and mice?
Does a simplified post-processing technique accurately quantify cardiac mechanics compared to the original method in humans and mice?
A simplified post-processing technique for DENSE CMR reduces processing time by 94% while maintaining high accuracy for quantifying cardiac mechanics, potentially facilitating its clinical adoption.
Answers are grounded in this paper and cite their sources.
May streamline DENSE CMR research workflows; leaves open prospective validation before clinical adoption.
BACKGROUND: Cardiovascular magnetic resonance using displacement encoding with stimulated echoes (DENSE) is capable of assessing advanced measures of cardiac mechanics such as strain and torsion. A potential hurdle to widespread clinical adoption of DENSE is the time required to manually segment the myocardium during post-processing of the images. To overcome this hurdle, we proposed a radical approach in which only three contours per image slice are required for post-processing (instead of the typical 30-40 contours per image slice). We hypothesized that peak left ventricular circumferential, longitudinal and radial strains and torsion could be accurately quantified using this simplified analysis. METHODS AND RESULTS: We tested our hypothesis on a large multi-institutional dataset consisting of 541 DENSE image slices from 135 mice and 234 DENSE image slices from 62 humans. We compared measures of cardiac mechanics derived from the simplified post-processing to those derived from original post-processing utilizing the full set of 30-40 manually-defined contours per image slice. Accuracy was assessed with Bland-Altman limits of agreement and summarized with a modified coefficient of variation. The simplified technique showed high accuracy with all coefficients of variation less than 10% in humans and 6% in mice. The accuracy of the simplified technique was also superior to two previously published semi-automated analysis techniques for DENSE post-processing. CONCLUSIONS: Accurate measures of cardiac mechanics can be derived from DENSE cardiac magnetic resonance in both humans and mice using a simplified technique to reduce post-processing time by approximately 94%. These findings demonstrate that quantifying cardiac mechanics from DENSE data is simple enough to be integrated into the clinical workflow.
No takes yet. Share an insight, caveat, or question.
Suever et al. (2014) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: