Key result
Warfarin concentration in oral fluid was strongly correlated with INR in patient samples with a pH of 7.2 or higher.
Why the study?
Does measurement of warfarin in oral fluid correlate with INR in patients on warfarin therapy?
Cross-Sectional (n=50)
No
Does measurement of warfarin in oral fluid correlate with INR in patients on warfarin therapy?
Warfarin can be measured in oral fluid and correlates with INR when oral fluid pH is ≥7.2, suggesting a potential non-invasive monitoring method.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for non-invasive warfarin monitoring; prospective validation required before any clinical adoption.
BACKGROUND: Patients on warfarin therapy undergo invasive and expensive checks for the coagulability of their blood. No information on coagulation levels is currently available between two controls. METHODOLOGY: A method was developed to determine warfarin in oral fluid by HPLC and fluorimetric detection. The chromatographic separation was performed at room temperature on a C-18 reversed-phase column, 65% PBS and 35% methanol mobile phase, flow rate 0.7 mL/min, injection volume 25 µL, excitation wavelength 310 nm, emission wavelength 400 nm. FINDINGS: The method was free from interference and matrix effect, linear in the range 0.2-100 ng/mL, with a detection limit of 0.2 ng/mL. Its coefficient of variation was <3% for intra-day measurements and <5% for inter-day measurements. The average concentration of warfarin in the oral fluid of 50 patients was 2.5±1.6 ng/mL (range 0.8-7.6 ng/mL). Dosage was not correlated to INR (r = -0.03, p = 0.85) but positively correlated to warfarin concentration in the oral fluid (r = 0.39, p = 0.006). The correlation between warfarin concentration and pH in the oral fluid (r = 0.37, p = 0.009) confirmed the importance of pH in regulating the drug transfer from blood. A correlation between warfarin concentration in the oral fluid and INR was only found in samples with pH values ≥7.2 (r = 0.84, p = 0.004). CONCLUSIONS: Warfarin diffuses from blood to oral fluid. The method allows to measure its concentration in this matrix and to analyze correlations with INR and other parameters.
No takes yet. Share an insight, caveat, or question.
Ghimenti et al. (2011) conducted a cross-sectional in Patients on warfarin therapy (n=50). Warfarin concentration in oral fluid was evaluated on Correlation between warfarin concentration in oral fluid and INR in samples with pH ≥ 7.2 (r = 0.84, p=0.004). Warfarin concentration in oral fluid was strongly correlated with INR in patient samples with a pH of 7.2 or higher.