Key result
Right atrial His-bundle pacing proves feasible for achieving QRS narrowing in prosthetic tricuspid valves.
Why the study?
Tricuspid valve prosthesis presents an anatomical barrier to right ventricular pacing, raising the question of whether permanent His-bundle pacing from the right atrium is feasible.
Is permanent His-bundle pacing from the right atrium feasible in patients with a prosthetic tricuspid valve and complete atrioventricular block?
Case Report (n=2)
Is permanent His-bundle pacing from the right atrium feasible in patients with a prosthetic tricuspid valve and complete atrioventricular block?
Answers are grounded in this paper and cite their sources.
Demonstrates that permanent His-bundle pacing from the right atrial side of the tricuspid annulus is a feasible physiological pacing strategy for patients with mechanical or bioprosthetic tricuspid valves where crossing the valve into the right ventricle is contraindicated.
Key Teaching Points•Tricuspid valve prosthesis is an anatomical barrier to right ventricular pacing.•The anatomical location of optimal His-bundle pacing site has been reported to be in the right atrium (RA) side of the tricuspid valve.•The RA portion of the His bundle remains anatomically unaltered by tricuspid valve prosthesis; hence, permanent His-bundle pacing from the RA is a feasible option for patients with inaccessible right ventricle. •Tricuspid valve prosthesis is an anatomical barrier to right ventricular pacing.•The anatomical location of optimal His-bundle pacing site has been reported to be in the right atrium (RA) side of the tricuspid valve.•The RA portion of the His bundle remains anatomically unaltered by tricuspid valve prosthesis; hence, permanent His-bundle pacing from the RA is a feasible option for patients with inaccessible right ventricle.
No takes yet. Share an insight, caveat, or question.
Rojas et al. (2019) conducted a case report in Complete atrioventricular block following tricuspid valve replacement (n=2). Permanent His-bundle pacing from the right atrium with backup coronary sinus left ventricular pacing lead was evaluated on Feasibility of selective or para-Hisian His-bundle capture and QRS narrowing from the right atrium.
Synapse has enriched 10 related papers. Consider them for comparative context: