Differential pulse polarography was utilized in voltammetric immunoassay to determine human albumin and albumin antiserum. Albumin was labelled with lead, which exhibited a reduction peak with E1/2 at −680 mV vs SCE. The current is proportional to the concentration of lead labelled albumin, and hence the albumin at a fixed lead concentration, and it decreases in proportion to the amount of albumin antiserum added. Albumin concentrations of 2.0 × 10−6–10 × 10−6 M and 10–50 μl of goat albumin antiserum (titer 1:16 = maximum dilution causing precipitation with equal volume of 1 mg/ml human albumin) were measured in the presence of 2 × 10−6 M lead.