Answers are grounded in this paper and cite their sources.
A simple ion-pair-assisted in situ surfactant-based solid phase microextraction method was developed for the separation and preconcentration of trace Au(III) prior to flame atomic absorption spectrometric determination. The method combines the ion-association capability of tetrabutylammonium with in situ generation of a separable surfactant-based solid phase using sodium perchlorate. Under the optimized conditions, Au(III) extraction was performed at pH 4.0 in the presence of tetrabutylammonium bromide and NaClO4. The extraction phase was isolated by centrifugation and subsequently dissolved in ethanol for flame atomic absorption spectrometric measurement. The effects of pH, surfactant type and concentration, NaClO4 concentration, sample volume, ionic strength, extraction time, and centrifugation conditions were systematically investigated. Under optimized conditions, a linear analytical response was obtained for Au(III) concentrations of 1.5–130.0 µg L−1, with a coefficient of determination (R2) of 0.9977. The limit of detection was 0.42 µg L−1, and the repeatability was 3.1% based on ten replicate measurements. The theoretical preconcentration factor was 80. The method was applied to mineral, tap, river, and seawater samples, with recoveries of 97–102% for spiked samples. These results demonstrate the potential of the proposed approach for Au(III) separation and preconcentration in aqueous samples before flame atomic absorption spectrometric determination.
No takes yet. Share an insight, caveat, or question.
Jamali et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: