Key result
AI-based 3DE parameters significantly correlated with CMR (P<0.05) and identified RV dysfunction with 94% sensitivity and 67% specificity at an RVEF cutoff of 43%.
Why the study?
The validation and diagnostic value of multiple right ventricle volumes and functional parameters derived from a novel AI-based 3DE algorithm needed exploration in comparison with CMR.
Does AI-based 3DE accurately quantify right ventricular size and function compared to cardiac magnetic resonance in patients with various clinical diagnoses?
Observational (n=51)
Does AI-based 3DE accurately quantify right ventricular size and function compared to cardiac magnetic resonance in patients with various clinical diagnoses?
p-value: p=<0.05
AI-based 3D echocardiography provides rapid and accurate quantification of right ventricular volumes and function, correlating well with the gold-standard cardiac magnetic resonance.
May support AI-3DE for RV assessment; leaves open prospective validation before clinical adoption.
Aims: This study aimed to explore the validation and the diagnostic value of multiple right ventricle (RV) volumes and functional parameters parameters derived from a novel artificial intelligence (AI)-based three-dimensional echocardiography (3DE) algorithm compared to cardiac magnetic resonance (CMR). Methods and Results: 51 patients with a broad spectrum of clinical diagnoses were finally included in this study. AI-based RV 3DE was performed in a single-beat HeartModel mode within 24 hours after CMR. Whether in the entire population or the patients with moderate and poor image quality, RV volumes and right ventricular ejection fraction (RVEF) measured by AI-based 3DE showed a statistically significant correlation with the corresponding CMR analysis (P<0.05 for all). The Bland-Altman plots indicated that these parameters were slightly underestimated by AI-based 3DE. Based on CMR derived RVEF<45% as RV dysfunction, end-systolic volume (ESV), end-systolic volume index (EDVi), stroke volume (SV), and RVEF showed great diagnostic performance in identifying RV dysfunction, as well as some non-volumetric parameters, including tricuspid annular systolic excursion (TAPSE), fractional area change (FAC), RV septum and free-wall longitudinal strains (LS) (P<0.05 for all). The cutoff value was 43% for RVEF with a sensitivity of 94% and specificity of 67%. Conclusion: AI-based 3DE provide rapid and accurate quantitation of the RV volumes and function with multiple parameters. Both volumetric and non-volumetric measurements derived from AI-based 3DE contributed to the identification of the RV dysfunction, even in the patients without excellent image quality of RV 3DE.
No takes yet. Share an insight, caveat, or question.
Zhu et al. (2021) conducted an observational in broad spectrum of clinical diagnoses (n=51). Artificial intelligence (AI)-based three-dimensional echocardiography (3DE) vs. Cardiac magnetic resonance (CMR) was evaluated on Correlation of RV volumes and RVEF with CMR analysis (p=<0.05). AI-based 3DE parameters significantly correlated with CMR (P<0.05) and identified RV dysfunction with 94% sensitivity and 67% specificity at an RVEF cutoff of 43%.