Why the study?
Do wrist blood pressure-measuring devices (Omron RX and Nissei WS-310) provide accurate blood pressure readings compared to a standard mercury manometer in hypertensive outpatients?
Population
87 patients selected from 152 subjects attending an out-patient hypertensive clinic according to AAMI/BHS…
Comparison
Blood pressure measurement using wrist… vs Standard auscultatory method using a mercury…
Design
Cross-sectional
Loading...
Wrist devices may aid diastolic BP monitoring in outpatients; leaves open systolic reliability versus mercury standards.
Do wrist blood pressure-measuring devices (Omron RX and Nissei WS-310) provide accurate blood pressure readings compared to a standard mercury manometer in hypertensive outpatients?
Wrist oscillometric devices (Omron RX and Nissei WS-310) are accurate for measuring diastolic blood pressure, but show decreased reliability for systolic blood pressure compared to standard mercury sphygmomanometers.
A 2002 study studied this question.