Does a wrist cuff oscillometric blood pressure device (OMRON R3) accurately measure blood pressure compared to intraarterial and mercury manometer measurements in patients undergoing cardiac catheterization?
Population
100 consecutive patients undergoing cardiac catheterization
Comparison
Wrist cuff oscillometric blood pressure… vs Intraarterial measurement and arm cuff mercury…
Design
Cross-sectional, Measurements taken in random order
Authors
Loading...
Wrist cuff may offer reliable readings in catheterization; leaves open routine adoption versus mercury manometer.
Does a wrist cuff oscillometric blood pressure device (OMRON R3) accurately measure blood pressure compared to intraarterial and mercury manometer measurements in patients undergoing cardiac catheterization?
The wrist cuff oscillometric device measures arterial blood pressure with high accuracy and reproducibility, showing better correlation with intraarterial diastolic pressure than a mercury manometer.