January 1, 1999Kidney & Blood Pressure Research

Wrist Measurement of Blood Pressure: Some Critical Remarks to Oscillometry

View Full Paper
Ask AI
Bookmark
Share

Why the study?

Does wrist oscillometric blood pressure measurement accurately reflect invasive aortic blood pressure in patients undergoing coronary angiography?

Population

27 subjects at the end of a coronary angiography procedure

Comparison

Wrist oscillometric blood pressure measurement… vs Invasive aortic arch blood pressure measurement…

Design

Cross-sectional

Authors

FWF. WeberRoche (Switzerland)RERaimund ErbelInterventional CardiologyRSRainhild SchäfersUniversity of Münster

Discussion

Loading...

Member takes

Implication

Wrist devices vary substantially from aortic BP; leaves open their clinical reliability pending rigorous validation.

Key Points

  • This research aims to evaluate the accuracy of wrist oscillometric blood pressure measurement devices compared to a standard method.
  • Measured blood pressure at the aortic arch and wrist in 27 subjects during coronary angiography.
  • Performed four replicate comparative measurements for reliability.
  • Analyzed correlation coefficients and differences, applying multiple regression for diastolic pressure.
  • The BPW measured systolic pressure 1.2 mm Hg higher and diastolic pressure 4.1 mm Hg higher on average compared to standard measurement.
  • Correlation coefficients were 0.85 for systolic and 0.84 for diastolic pressures.
  • Significant variability in measurements was found, with poorer accuracy for lower diastolic values (r = 0.89, p<0.001).

Structured PICO

Does wrist oscillometric blood pressure measurement accurately reflect invasive aortic blood pressure in patients undergoing coronary angiography?

P
Population
27 subjects at the end of a coronary angiography procedure
I
Intervention
Wrist oscillometric blood pressure measurement using the blood pressure watch (BPW) by NAIS-Matsushita
C
Comparator
Invasive aortic arch blood pressure measurement using a Statham P23 transducer
O
Outcome
Accuracy of systolic and diastolic blood pressure measurement (differences between methods)surrogate

Wrist oscillometric blood pressure devices show significant variability and reduced accuracy at lower diastolic pressures compared to invasive aortic measurements, highlighting the need for rigorous validation.

Cite This Study

Weber et al. (1999) studied this question.

synapsesocial.com/papers/6a72c4138031ec7bb1dd7bd7https://doi.org/10.1159/000025923
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Bias and variability in blood pressure measurement with ambulatory recorders.1993 · 43 citations
  2. 2Arterial Blood Pressure Measurements with a Portable Recorder in Hypertensive Patients1964 · 149 citations
  3. 3Automated Ambulatory Blood Pressure Devices and Self-Measured Blood Pressure Monitoring Devices: Their Role in the Diagnosis and Management of Hypertension1993 · 58 citations
  4. 4Ambulatory blood pressure monitoring in normotensive children1994 · 110 citations
  5. 5The Prognostic Value of Ambulatory Blood Pressures1983 · 856 citations