Key result
An automated wrist device systematically underestimated systolic (120.1 vs 130.5 mmHg, P<0.001) and diastolic BP compared with an arm device in elderly persons.
Why the study?
Does an automated oscillometric wrist device with position sensor provide accurate blood pressure measurements compared to an arm device in elderly persons?
RCT (n=48)
random order
Does an automated oscillometric wrist device with position sensor provide accurate blood pressure measurements compared to an arm device in elderly persons?
Wrist blood pressure devices with position sensors systematically underestimate systolic and diastolic blood pressure compared to arm devices in elderly persons, questioning their clinical utility in this population.
Answers are grounded in this paper and cite their sources.
Wrist devices may underestimate BP in elderly patients; supports caution in use and leaves open need for prospective validation.
OBJECTIVE: Home blood pressure (BP) monitoring is recommended by several clinical guidelines and has been shown to be feasible in elderly persons. Wrist manometers have recently been proposed for such home BP measurement, but their accuracy has not been previously assessed in elderly patients. METHODS: Forty-eight participants (33 women and 15 men, mean age 81.3±8.0 years) had their BP measured with a wrist device with position sensor and an arm device in random order in a sitting position. RESULTS: Average BP measurements were consistently lower with the wrist than arm device for systolic BP (120.1±2.2 vs. 130.5±2.2 mmHg, P<0.001, means±SD) and diastolic BP (66.0±1.3 vs. 69.7±1.3 mmHg, P<0.001). Moreover, a 10 mmHg or greater difference between the arm and wrist device was observed in 54.2 and 18.8% of systolic and diastolic measures, respectively. CONCLUSION: Compared with the arm device, the wrist device with position sensor systematically underestimated systolic as well as diastolic BP. The magnitude of the difference is clinically significant and questions the use of the wrist device to monitor BP in elderly persons. This study points to the need to validate BP measuring devices in all age groups, including in elderly persons.
No takes yet. Share an insight, caveat, or question.
Guggiari et al. (2013) conducted an RCT in Elderly persons (n=48). Automated oscillometric wrist device with position sensor vs. Automated oscillometric arm device was evaluated on Systolic blood pressure (p=<0.001). An automated wrist device systematically underestimated systolic (120.1 vs 130.5 mmHg, P<0.001) and diastolic BP compared with an arm device in elderly persons.
Synapse has enriched 5 related papers. Consider them for comparative context: