Key result
The OMRON R6 wrist-cuff device produced mean blood pressures (134/77 mmHg) comparable to the OMRON HEM 705 CP upper arm-cuff device in a horizontal position (137/80 mmHg).
Why the study?
Does the OMRON R6 wrist-cuff oscillometric device provide accurate blood pressure measurements compared to the OMRON HEM 705 CP upper arm-cuff device when arm position is controlled?
Cross-Sectional
Does the OMRON R6 wrist-cuff oscillometric device provide accurate blood pressure measurements compared to the OMRON HEM 705 CP upper arm-cuff device when arm position is controlled?
Wrist-cuff blood pressure monitors can provide accurate readings comparable to upper arm-cuff devices if the arm is properly positioned at heart level.
Answers are grounded in this paper and cite their sources.
Wrist-cuff BP readings vary with arm position; leaves open their clinical reliability versus validated arm-cuff devices.
BACKGROUND: Despite the increasing popularity of wrist-cuff blood pressure (BP) devices, their accuracy has not been established and international guidelines do not support their use. Because arm position influences BP measurement, it is possible that conflicting reports on wrist-cuff device accuracy reflects diverse arm positions. METHOD: This study compared BP measured by two oscillometric devices, the upper arm-cuff OMRON HEM 705 CP and the OMRON R6 oscillometric wrist-cuff device. In the former BP was measured with the arm in two supported positions, dependent on a table (manufacturer's instructions) and horizontal (mid sternum), while the latter followed the manufacturer's instructions. RESULTS: In contrast to the dependent arm where BP was significantly higher (P<0.05), the horizontal arm position with the arm-cuff produced a mean systolic and diastolic BP comparable to the wrist-cuff device where the wrist was at heart level being respectively, 137+/-29/80+/-16 and 134+/-27/77+/-16 mmHg. A close relationship over a wide BP range was also confirmed by least squares, least product linear regression and Bland-Altman analysis. CONCLUSION: This study supports the use of wrist-cuff monitors for self/home use and underlines the need for a more precise definition for arm position when using all BP devices -- mercury and oscillometric.
No takes yet. Share an insight, caveat, or question.
Mourad et al. (2005) reported a cross-sectional. OMRON R6 wrist-cuff device vs. OMRON HEM 705 CP upper arm-cuff device was evaluated on Mean systolic and diastolic blood pressure. The OMRON R6 wrist-cuff device produced mean blood pressures (134/77 mmHg) comparable to the OMRON HEM 705 CP upper arm-cuff device in a horizontal position (137/80 mmHg).