Why the study?
Wearable sensor nodes in wireless body area networks generate and transmit large amounts of biomedical data, driving up transmission power consumption and data storage needs during long-term monitoring.
The proposed VLSI-based hybrid ECG compression scheme offers a low-power and area-efficient solution for wearable sensor nodes in wireless body area networks.
Answers are grounded in this paper and cite their sources.
May facilitate extended wearable ECG monitoring; leaves open clinical validation in body-area networks.
During smart long-term monitoring of any biomedical signal in wireless body area networks, wearable sensor nodes generate and transmit a large amount of data, increasing transmission power consumption. In order to reduce data storage and power consumption, a lossless data compression technique for an electrocardiogram signal monitoring system is presented in this letter. For this, a hybrid lossless compression algorithm based on Run-length coding and Golomb–Rice coding is proposed to enhance the bit compressing rate. The lossless encoding scheme is implemented on the MIT-BIH arrhythmia database, achieving a compression ratio of 2.91. A VLSI-based architecture of the data compression algorithm is implemented in 90nm CMOS technology that consumes power of 18.78μ Wat 100 MHz operating frequency and 1.2 V supply voltage, occupying an area of 0.0051mm².