To reduce the volume of linear totally implantable artificial hearts, the drive velocity of linear pulse motors must be increased, potentially through application to nonpulsatile artificial hearts.
Answers are grounded in this paper and cite their sources.
Brushless DC motors may enable smaller implantable hearts; leaves open LPM viability via velocity optimization for nonpulsatile designs.
The authors have been considering the possibility of using a linear pulse motor (LPM) as a driving source, and have developed a series of artificial heart models that incorporate such a motor. The newly developed linear-type totally implantable artificial heart (linear TAH) has two blood pumps, which are inflated and deflated alternately by the LPM, causing a pusher plate to pulsate the blood. This paper compares the machine parameters of LPMs, brush- less dc motors, and ultrasonic motor for driving artificial hearts. We selected two parameters for the comparison: the mechanical output/volume ratio and the motor constant. The results may be summarized as follows: (1)the brushless dc motors had a mechanical output/volume ratio 3 to 5.5 times larger than that of the LPMs; (2) we considered that the motor constant of the LPM was within almost the same range as that of the brushless dc motors. It was established that in order to reduce the volume of the linear TAH, it is necessary to increase the drive velocity of the LPMs by using a special mechanism. We consider that application of the LPM to a nonpulsatile artificial heart is one way to achieve this.
No takes yet. Share an insight, caveat, or question.
Yamaguchi et al. (1996) studied this question.