Key result
A linear motor-driven total artificial heart was developed and evaluated for pump performance in vitro and in an acute in vivo animal test using a sheep.
The study describes the initial development and preclinical testing (in vitro and in vivo in a sheep) of a novel linear motor-driven total artificial heart.
Answers are grounded in this paper and cite their sources.
Initial preclinical data support further development of linear motor-driven total artificial hearts; leaves open clinical translation.
The authors first explain the structure and features of the linear pulse motor (LPM) developed to drive the total artificial heart (TAH). Then they describe the linear motor-driven TAH using the LPM as the driving source, which the authors call linear TAH, and the configuration of a human model in which the linear LAH is installed. Next, the authors describe the results of evaluating the pump performance of the linear TAH in vitro by using a mechanical mock circulatory testing system. Finally, the authors report information resulting from the evaluation of the linear TAH in vivo in an acute animal test using a sheep.>
No takes yet. Share an insight, caveat, or question.
Yamada et al. (1995) studied this question. Linear motor-driven total artificial heart (linear TAH) was evaluated on Pump performance in vitro and in vivo. A linear motor-driven total artificial heart was developed and evaluated for pump performance in vitro and in an acute in vivo animal test using a sheep.
Synapse has enriched 10 related papers. Consider them for comparative context: