An automated colony counter was found to readily detect surface and subsurface bacterial colonies of 0.3-mm size or greater with a high degree of precision. On a logarithmic scale, counting efficiency consistently ranged from 89 to 95% of corresponding manual count determinations for plates containing up to 1,000 colonies. In routine application, however, automated plate counts up to approximately 400 colonies were selected as a more practical range for operation. The automated counter was easily interfaced with an automated data acquisition system.
No takes yet. Share an insight, caveat, or question.
Goss et al. (1974) studied this question.