Answers are grounded in this paper and cite their sources.
It is supposed that a commercial cavity ionization chamber fitted with its Perspex build-up cap, as required for a 2 MV X-ray exposure calibration, is centred at a specified depth in a tissue-like phantom for a dose determination. The source of radiation may be an X-ray source in the range 2 MV to 15 MV X-rays or a γ-ray source 137 Cs or 60 Co. However, the scale of the associated measuring instrument is determined by the 2 MV X-ray exposure calibration. Hence, a reading in units of 2 MV X-ray ` roentgens ' is always given, irrespective of the situation of the chamber and the source of radiation used with the phantom. The problem presented is how to convert a scale reading in units of 2 MV X-ray ` roentgens ' to a scale reading in rads for a particular phantom material and a particular X- or γ-ray source. These computational problems are examined in the paper. An expression is derived for the required conversion factor. It is shown that these conversion factors are not critically dependent on the composition of the medium for low- Z materials or on small changes in the dimensions of the cavity chambers. The calculations refer mainly to a typical 0·6 cm 3 Baldwin-Farmer chamber; but some results are given also for a Victrometer chamber.
No takes yet. Share an insight, caveat, or question.
G P Barnard (1964) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: