OBJECTIVE: study, which simulated ablation of a 2.5-cm virtual hepatic tumor. MATERIALS AND METHODS: A total of 132 ablation zones were created (22 ablation zones for each protocol) using three separable clustered electrodes. The performances of the DSM, SB, and combined DSM + SB ablation modes were compared by evaluating the following parameters of the RFA zones at two interelectrode distances: shape (circularity), size (diameter and volume), peritumoral ablative margins, and percentages of the white zone at the midpoint of the two electrodes (ablative margin at midpoint, AMm) and in the electrode path (ablative margin at electrode path, AMe). RESULTS: = 0.002 for 25-mm and 20-mm, respectively). All size measurements, AMm, and AMe were the greatest in the DSM mode, followed by the DSM + SB mode, and the lowest were with the SB mode (all statistically significant). The white zone proportion in AMm and AMe were the greatest in the SB mode, followed by the DSM + SB mode and DSM in general. CONCLUSION: DSM and SB appear to be complementary in creating an ideal ablation zone. RFA with the SB mode can efficiently eradicate tumors and create a circular ablation zone, while DSM is required to create a sufficient ablative margin and a large ablation zone.