Energy dispersive X-ray fluorescence (ED-XRF) spectrometer was built in the laboratory with low wattage X-ray tube and silicon drift detector using a 3D printer. SUS 304 and 316L stainless steels were measured by the ED-XRF spectrometer and a sequential type inductively-coupled plasma atomic emission spectrometer (ICP-AES). The results of quantitative analysis obtained by the fundamental parameter (FP) program developed by the authors were compared with the ICP-AES results, and concluded the following three points.