Key result
Pannus overgrowth was the primary cause of aortic St. Jude Medical valve obstruction, accounting for 7 of 9 redo valve replacement events in patients on adequate oral anticoagulation.
Observational (n=8)
Pannus overgrowth appears to be the primary mechanism for obstruction of aortic St. Jude Medical valves, even in patients on adequate oral anticoagulation.
Answers are grounded in this paper and cite their sources.
Consider pannus in obstructed aortic mechanical valves despite anticoagulation; case report data leave mechanism and management questions open.
Between 1995 and 2000, 8 patients with St. Jude Medical (SJM) valves in the aortic position required 9 redo valve replacement for prosthetic valve obstruction. Obstruction of the prosthetic valve was diagnosed by simultaneous echocardiography and cineradiography, and process of restricted leaflet movement that progressed to hemodynamic impairment was observed by serial studies in three recent patients. An oral anticoagulation was considered to be adequate in all patients except one patient who had withdrawal of warfrain. Pannus was the sole cause of valve obstruction in seven events in 6 patients, and both thrombus and pannus in 2 patients. Pannus overgrowth was found on the inflow aspect of the SJM valve, and involved the ends of the straight edge of the leaflets over pivot guards. These results suggest that pannus might play the primary role in development of obstruction of aortic SJM valves in patients on adequate oral anticoagulation.
No takes yet. Share an insight, caveat, or question.
S Aoyagi (2002) conducted an observational in Prosthetic valve obstruction (n=8). St. Jude Medical valves in the aortic position was evaluated on Cause of prosthetic valve obstruction. Pannus overgrowth was the primary cause of aortic St. Jude Medical valve obstruction, accounting for 7 of 9 redo valve replacement events in patients on adequate oral anticoagulation.
Synapse has enriched 10 related papers. Consider them for comparative context: