Answers are grounded in this paper and cite their sources.
OBJECTIVES: The aim of this work was to implement a fast, accurate and simple method to quantify plasma ADMA and SDMA, in a run time suitable for routine analysis. DESIGN AND METHODS: We developed and validated a hydrophilic interaction chromatographic method coupled to tandem mass spectrometry (HILIC-MS/MS) for separation and simultaneous quantification of Arginine (Arg) and its dimethylarginines, ADMA and SDMA, with a short run time (less than 5 min) using a small volume of human plasma (0.02 mL). RESULTS: Correlation coefficients (r) of the calibration curves ranged from 0.9926 to 0.9984. Within-day and between-day imprecision (CV%) and inaccuracy (%), carry-over and recovery were also evaluated for validation. Preliminary data of Arg, ADMA and SDMA from 30 apparently healthy subjects and type 2 diabetic patients (n=33) with and without kidney dysfunction were calculated and some statistical differences occurred among them (p<0.05). CONCLUSIONS: Data from calibration curves and quality controls reveal that the method is accurate and precise. Healthy subjects and diabetic patients' values are in agreement with those reported in other studies.
No takes yet. Share an insight, caveat, or question.
D’Apolito et al. (2008) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: