Why the study?
Ballistocardiogram signals from optical fiber sensors are difficult to detect due to differential detector effects, self-interference, installation variation, individual differences, and body interferences.
Does an optical fiber sensor mattress with adaptive regulations accurately estimate heart and respiration rates compared to conventional medical monitors in healthy volunteers?
Does an optical fiber sensor mattress with adaptive regulations accurately estimate heart and respiration rates compared to conventional medical monitors in healthy volunteers?
An optical fiber sensor integrated into a mattress can accurately estimate heart and respiration rates under various posture conditions without direct skin contact.
Answers are grounded in this paper and cite their sources.
Supports the feasibility of contactless vital sign monitoring via optical fiber mattresses; leaves open whether this accuracy persists in.
The aim of this work is to present a method for accurately estimating heart and respiration rates under different actual conditions based on a mattress which was integrated with an optical fiber sensor. During the estimation, a ballistocardiogram (BCG) signal, which was obtained from the optical fiber sensor, was used for extracting the heart rate and the respiration rate. However, due to the detrimental effects of the differential detector, self-interference, and variation of installation status of the sensor, the ballistocardiogram (BCG) signal was difficult to detect. In order to resolve the potential concerns of individual differences and body interferences, adaptive regulations and statistical classifications spectrum analysis were used in this paper. Experiments were carried out to quantify heart and respiration rates of healthy volunteers under different breathing and posture conditions. From the experimental results, it could be concluded that (1) the heart rates of 40-150 beats per minute (bpm) and respiration rates of 10-20 breaths per minute (bpm) were measured for individual differences; (2) for the same individuals under four different posture contacts, the mean errors of heart rates were separately 1.60 ± 0.98 bpm, 1.94 ± 0.83 bpm, 1.24 ± 0.59 bpm, and 1.06 ± 0.62 bpm, in contrast, the mean errors of the polar beat device were 1.09 ± 0.96 bpm, 1.44 ± 0.99 bpm, and 1.78 ± 0.94 bpm. Furthermore, the experimental results were validated by conventional counterparts which used skin-contacting electrodes as their measurements. It was reported that the heart rate was 0.26 ± 2.80 bpm in 95% confidence intervals (± 1.96SD) in comparison with Philips sure-signs VM6 medical monitor, and the respiration rate was 0.41 ± 1.49 bpm in 95% confidence intervals (± 1.96SD) in comparison with ECG-derived respiratory (EDR) measurements for respiration rates. It was indicated that the developed system using adaptive regulations and statistical classifications spectrum analysis performed better and could easily be used under complex environments.