This is the authors' abstract. We don't add key points for this paper.
In this article, we describe the design rationale and performance features of an integrated multichannel continuous wave (cw) near-infrared (NIR) optical tomographic imager capable of collecting fast tomographic measurements over a large dynamic range. Fast data collection (∼70 Hz/channel/wavelength) is achieved using time multiplexed source illumination (up to 25 illumination sites) combined with frequency encoded wavelength discrimination (up to four-wavelength capability) and parallel detection (32 detectors). The described system features a computerized user interface that allows for automated system operation and is compatible with various previously described measuring heads. The results presented show that the system exhibits a linear response over the full dynamic measuring range (180 dB), and has excellent noise (∼10 pW noise equivalent power) and stability performance (<1% over 30 min). Recovered images of laboratory vessels show that dynamic behavior can be accurately defined and spatially localized.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Schmitz et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: