Why the study?
Does two-dimensional echocardiography accurately estimate volume in asymmetric model hearts compared to direct measurement?
Does two-dimensional echocardiography accurately estimate volume in asymmetric model hearts compared to direct measurement?
Biplane two-dimensional echocardiography using the disc method provides accurate volume determination in asymmetric model hearts, which has implications for left ventricular volume assessment in patients with coronary artery disease.
Answers are grounded in this paper and cite their sources.
Biplane disc method may aid asymmetric LV volumetry in models; leaves open validation for clinical CAD assessment.
In asymmetric model hearts volume was estimated by two-dimensional echocardiography (2-dE) and radiography, and the results compared to directly measured volume in the range from 10 to 320 ml (n = 22). The disc method, and the area-length and ellipsoid methods were used for both single-plane and biplane calculations. When biplane 2-dE measurements were calculated by the disc method and compared with directly measured volumes the regression equation was acceptably linear. For single-plane calculations, no linearity was found. For the disc method, the regression equation was given by y = 0.91x + 8.4, the correlation coefficient (r) was 0.989, and the standard error of estimate (s.e.e.) was ± 12.8 ml. The confidence interval was 0.85 to 0.98. With the area-length method the regression equation was y = 0.93x + 7.4, r = 0.987, s.e.e. = ± 14.0 ml. For the ellipsoid method, the regression equation was y = 0.89x + 1.2, r = 0.966, s.e.e. = ±22.0 ml. With radiographic biplane calculations compared to directly determined values the regression equation was also found to be linear. For single-plane calculations such linearity was not found. For the disc method the regression equation was y = 0.99x + 5.9, r = 0.991, and s.e.e. = ±12.6 ml. The confidence interval was 0.93 to 1.06. For the area-length method the regression equation was y = 1.01x + 9.0, r = 0.987, s.e.e. = ±15.0 ml, and for the ellipsoid method y = 0.97x − 3.3, r = 0.992, s.e.e. = ±11.2 ml. Thus, 2-dE is an accurate method for volume determination in asymmetric model hearts using biplane calculations and the disc method. Area-length and ellipsoid methods seemed to be less accurate. Compared to radiography, 2-dE yielded similar correlation coefficients and s.e.e.s. Whereas 2-dE slightly underestimated model heart volume (−4.5 ± 14.7 ml, NS), radiography overestimated the volume (+5.7 ± 12.3 ml, P<0.05). The mean difference between 2-dE and radiography was 10.2 ± 21.2 ml, P<0.05). These results ought to be taken into account when left ventricular volume determinations are done in patients with coronary artery disease.
No takes yet. Share an insight, caveat, or question.
Erbel et al. (1982) studied this question.