Key result
Non-invasive stroke volume measurement using bioreactance and pulse wave transit time showed unacceptable percentage errors and are not interchangeable with esophageal Doppler monitoring.
Why the study?
The study evaluated whether non-invasive stroke volume measurements using bioreactance or pulse wave transit time are interchangeable with esophageal Doppler monitoring during surgery.
Does non-invasive stroke volume measurement using bioreactance or pulse wave transit time provide interchangeable accuracy and trending ability compared to esophageal Doppler monitoring in adults undergoing gastrointestinal surgery?
Observational (n=10)
No
Does non-invasive stroke volume measurement using bioreactance or pulse wave transit time provide interchangeable accuracy and trending ability compared to esophageal Doppler monitoring in adults undergoing gastrointestinal surgery?
Effect estimate: correlation coefficient 0.75 (95% CI 0.62 to 0.78)
p-value: p=<0.001
Non-invasive stroke volume measurements using bioreactance and pulse wave transit time showed unacceptable percentage errors compared to esophageal Doppler monitoring, indicating they are not interchangeable.
May aid non-invasive SV monitoring in surgery; leaves open accuracy versus EDM pending larger validation.
Objective: To compare the trending ability, accuracy, and precision of non-invasive stroke volume (SV) measurement based on a bioreactance technique and measurement of the pulse wave transit time (PWTT) versus the esophageal Doppler monitoring (EDM). Materials and Methods: Two hundred twenty-seven paired measurements from 10 patients who underwent abdominal surgery under general anesthesia were included for SV measurements. Pearson’s correlation coefficient was calculated, and Bland-Altman analysis was performed to evaluate the agreement between EDM and bioreactance (EDM-bioreactance) and between EDM and PWTT (EDM-PWTT). Results: EDM-bioreactance had a correlation coefficient of 0.75 (95% confidence interval [CI] 0.62 to 0.78; p<0.001), bias of 0.28 ml (limits of agreement –30.92 to 31.38 ml), and percentage error of 46.82%. EDM-PWTT had a correlation coefficient of 0.48 (95% CI 0.44 to 0.72; p<0.001), bias of –0.18 ml (limits of agreement –40.28 to 39.92 ml), and percentage error of 60.17%. A subgroup analysis of data from patients who underwent crystalloid loading was performed to detect the trending ability. The four-quadrant plot analysis between EDM-bioreactance and EDM-PWTT demonstrated concordance rates of 70.00% and 73.68%, respectively. Conclusion: SV measurement based on bioreactance technique and measurement of PWTT are not interchangeable with EDM. Trial registration: Thai Clinical Trials Registry, TCTR 20181217003 Keywords: Stroke volume, Cardiac output, Doppler, Perioperative care, Pulse, Time
No takes yet. Share an insight, caveat, or question.
A 2020 study conducted an observational in Gastrointestinal surgery under general anesthesia (n=10). Bioreactance and pulse wave transit time (PWTT) monitoring vs. Esophageal Doppler monitoring (EDM) was evaluated on Correlation of stroke volume measurement between EDM and bioreactance (correlation coefficient 0.75, 95% CI 0.62 to 0.78, p=<0.001). Non-invasive stroke volume measurement using bioreactance and pulse wave transit time showed unacceptable percentage errors and are not interchangeable with esophageal Doppler monitoring.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: