Key result
M-mode echocardiography erroneously indicated 1:1 atrioventricular conduction in a neonate with atrial flutter and 2:1 conduction due to pseudo 1:1 movement of the atrial wall adjacent to the AV valve.
Case Report (n=1)
No
When using M-mode echocardiography to diagnose fetal or neonatal tachyarrhythmia, assessing the atrial wall far from the ventricle is necessary to avoid erroneous diagnosis of 1:1 AV conduction.
Answers are grounded in this paper and cite their sources.
Caution advised interpreting M-mode for AV conduction in neonatal flutter; leaves open validation of multi-modality protocols.
M-mode echocardiography has been playing an important role in the diagnosis of fetal tachyarrhythmia. We recently encountered a neonatal case of atrial flutter with 2:1 atrioventricular conduction. However, M-mode erroneously indicated 1:1 atrioventricular movement. While the movement of the atrial wall far from the atrioventricular valve was much faster than that of the ventricular wall, the atrial wall adjacent to the atrioventricular valve fully synchronized to that of the ventricular wall. Thus, to avoid this novel pitfall, it would be important to add an additional assessment focusing on the movement of the atrial wall far from the ventricle.
No takes yet. Share an insight, caveat, or question.
Kawano et al. (2018) conducted a case report in Neonatal atrial flutter (n=1). M-mode echocardiography was evaluated. M-mode echocardiography erroneously indicated 1:1 atrioventricular conduction in a neonate with atrial flutter and 2:1 conduction due to pseudo 1:1 movement of the atrial wall adjacent to the AV valve.